Water-in-Oil Monitors and Systems. Measurement of free Water in Oil. .
Range: | 0 – 600 ppm “free” water |
---|---|
Resolution: | 1 ppm, lower Range |
Accuracy: | Up to ± 10 ppm below 100 ppm |
Response Time: | < 5 sec. |
Linearity: | 0 – 100 ppm better than ± 2 % |
Display: | Full Colour Graphic |
Datalogging System | Yes |
Power Supply: | 24 V – 240 V AC or DC |
Consumption: | < 10 VA |
Alarm Points 1, 2: | Adjustable, (Works adjustment 70 ppm) |
Alarm 1 Operating Delay: (for annunciation purpose) | Adjustable between 1 – 540 sec. (Works adjustment 2 sec.) |
Alarm 2 Operating Delay: (for control purposes) | Adjustable between 1 – 540 sec. (Works adjustment 10 sec.) |
Alarm Indication: (Alarm 1, 2, System Fault) | Red visual indicator |
Alarm Contact Rating: (Alarm 1, 2, System Fault) | Potential free 1 pole change over contacts, 3 A / 240 V AC |
Output Signal: | 0 – 20 mA, ext. Load < 150 Ω Output range adjustable |
Sample Pressure: | 0 – 10 bar |
Sample Flow: | Approx. 2 l/min |
Sample Temperature: | + 1 to + 65° C |
Ambient Temperature: | + 1 to + 45° C |
Size: Computer Unit | 200 mm W x 200 mm H x 120 mm D |
Size: Measuring Cell | 130 mm W x 55 mm H x 70 mm D |
Weight: | 4 kg |
Degree of Protection: | IP 65 |
Pipe Connections: | R ¼" Female |
> Technical specifications are subject to change without notification. |